Composite digital membrane electrode for free chlorine, total chlorine or chlorine dioxide configurations

Free / Total Chlorine and Chlorine Dioxide Analyzer
The broadest documented disinfectant range in this five-model family, using a composite digital membrane electrode with no routine zero calibration and DPD-1 slope calibration.
The five products retain separate principles, ranges, sample limitations and interfaces. Final configuration is controlled by the project confirmation document.
Measurement method and product highlights
Special permeable membrane and antioxidant electrolyte documented for stable measurement
No routine zero calibration; slope calibration uses the DPD-1 reference method
Integrated automatic linear temperature compensation
Three 4–20 mA outputs, three configurable relays and Modbus RS485
Verified technical specifications
Confirm the disinfectant, required free or total chlorine result, normal and peak range, sample pH, conductivity, temperature, pressure, flow-cell conditions, calibration method and control interfaces before quotation.
- Measurement principle
- Composite digital membrane electrode
- Measurement ranges
- 0–2 mg/L; 0–20 mg/L; 0.5–200 mg/L
- Sample pH
- 4–9
- Conductivity range
- 10 µS/cm–50 mS/cm; seawater configuration documented
- Flow through cell
- 250–500 mL/min
- Slope drift
- Approx. -1% per month
- Response time
- T90 approx. 2 min
- Zero calibration
- Not required
- Slope calibration
- DPD-1 reference method
- Temperature compensation
- Integrated automatic linear
- Materials
- PVC-U, stainless steel and microporous hydrophilic membrane
- Power supply
- 85–500 VAC; 9–36 VDC
- Analog output
- 3 × 4–20 mA
- Relays
- 3 configurable relays
- Operating temperature
- Sample: 0–45°C, non-condensing; ambient: 0–55°C
- Communication
- Modbus RS485
- Sensor dimensions
- Ø25 × 205 mm
- Cable
- 5 m
Recommended process fit
Drinking-water treatment
Continuous disinfectant-residual monitoring at plant process and outlet points.
Distribution and secondary supply
Verify residual chlorine through distribution networks and secondary-supply facilities.
Cooling and process water
Support disinfectant control in circulating-water and industrial-water systems.
Wastewater disinfection
Track disinfectant residual at disinfection and discharge points when the selected method fits the water matrix.
Items to confirm before quotation
Confirm the target and water matrix before choosing DPD colorimetry, constant-voltage amperometry, micro-MEMS or a covered-membrane electrode.
- 01Disinfectant and free chlorine, total chlorine, chlorine dioxide or active-oxygen target
- 02Normal, peak and alarm range plus required detection limit
- 03pH, conductivity, temperature, pressure and sample cleanliness
- 04Flow-cell conditions, calibration, reagent or membrane maintenance
- 05Power, 4–20 mA, relays, Modbus, quantity and destination
Project documents available on request
After receiving the operating conditions, we prepare the configured datasheet, flow-cell and mounting confirmation, I/O schedule and quotation. The clean image is not a dimensional drawing.
Get a quote- Configured datasheet→
- Sample and mounting confirmation→
- Cable and I/O schedule→
- Project quotation→
Compare other chlorine models
Selection questions
Do all five analyzers use the same measurement method?+
No. CL-201 uses DPD colorimetry; PDCL-400 uses constant-voltage amperometry; PDCL-500M uses a micro-MEMS electrochemical structure; PDCL-700 and PDCL-600 use different membrane-electrode designs.
Can the model be selected only from the range?+
No. Confirm the disinfectant, free or total result, pH, conductivity, temperature, pressure, flow and maintenance method.
Can the product image be used as an installation drawing?+
No. The image provides consistent public presentation. Dimensions, interfaces, accessories and flow cells come from the datasheet and order confirmation.
Tell us what you need to measure.
We will respond with a recommended configuration, supporting documentation and quotation.